Liquiline System CA80AL offers precise, regulation-compliant online measurement of aluminum residues in drinking water and wastewater. The analyzer helps you to optimize aluminum dosing in flocculation processes and phosphate removal. It enables plug & play of up to 4 Memosens sensors – minimizing the installation effort. Automatic calibration and cleaning and the low reagent consumption reduce operating costs while advanced diagnostics with remote access ensure process safety.
Liquiline System CA80AL aluminum analyzer supports:
Limit monitoring in drinking water
Optimization of aluminum-based coagulation processes
Phosphate elimination in wastewater treatment plants
장점
The aluminum analyzer uses the pyrocatechol violet method according to DIN ISO 10566, ensuring direct comparability to lab results
Supports predictive maintenance by advanced diagnostics
Reduces operating costs thanks to automatic calibration and cleaning as well as low reagent consumption
Easy upgrade of functionality to a complete measuring station - by adding modules and connecting Memosens sensors.
Comfortable remote access via integrated web server
Seamless integration into process control systems thanks to digital field buses such as Modbus, PROFIBUS, EtherNet/IP
AVR Waste in Botlek, Rotterdam, processes up to 1.7 million tons of residual waste per year. Discover how they optimized their pH measurement thanks to Endress+Hauser
Using seawater to flush toilets can save a lot of drinking water. To ensure a constant water supply, the sensors used must function reliably to meet the water quality objectives (WQO).
Examination of wastewater, using the Liquistation CSF48 and the Liquiport CSP48 mobile, can provide valuable information about the occurrence of Covid 19 infections in a specific area
Flexibility and good ergonomics are particularly important in the production of heavy systems. Our modern production lines offer best possible support for our staff to deliver high quality products.
Process analyzer for aluminum in aqueous solutions Comply with standard colorimetric measuring principle - pyrocatechol violet methot
Size
Housing (open version): 793 x 530 x 417 mm 31.22 x 20.87 x 16.42 in Housing (closed version): 793 x 530 x 463 mm 31.22 x 20.87 x 18.23 in Housing with base: 1723 x 530 x 463 mm 67.83 x 20.87 x 18.23 in
Design
Open design, cabinet and stand housing High-Performance plastic ASA-PC, additional stand coated steel
Process temperature
4 ... 40 °C (39.2 ... 104 °F)
Ambient temperature
5 ... 40C ( 41 ... 104F) Outdoor version: - 20 to 40 °C (-4 to 104 °F)
Process pressure
at atmospheric pressure, < 0.2 bar
Sample flow rate
min. 5ml/min (0,17 fl.oz/min)
Consistency of the sample
low solid content (< 50NTU), aqueous homogeneous
Application
Monitoring of the aluminum content of potable water, mineral water and industrial waste water
2x 0/4 ... 20 mA Webserver, Modbus, Ethernet/IP, Profibus DP
Input
1 or 2 measuring channel 1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional)
Measuring range
15 to 1000 µg/l Al 15 to 1000 µg/l with dilution function to maximum 300 to 20 000 µg/l Al
Consumables
Reagents and standard solutions CY80AL as well as cleaner CY800 are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800.